KV-2 Low Temperature Viscometer Bath
- Mã sản phẩm: 94710-3
- Nhà sản xuất: Stanhope-Seta
- Giá :
Liên hệ đặt hàng:
KV-2 Low Temperature Viscometer Bath
Part Number 94710-3
The KV-2 Low Temperature Viscometer Bath is a compact low temperature high stability kinematic viscometer bath with integral refrigeration that can accommodate up to 2 viscometer tubes.
FEATURES
- Temperature range -40°C to +20°C
- High stability ±0.015 at -20°C, ±0.01 at -40°C
- Accommodates up to 2 standard viscometer tubes
- Illuminated Bath
- Insulated and heated viewing window
- Headed top plate
ADDITIONAL INFORMATION
The refrigeration system gives the stirred bath a temperature range of 20 down to -40°C, stable to within ±0.015 at -20°C. A twin display shows the current and set temperature. Temperature is measured by a PT100 probe. The bath has independent over-temperature and low fluid level cut-outs and a warning alarm.
The insulated bath is sealed by 8mm thick glass front and rear with 60mm plexiglass providing additional insulation. The bath is illuminated via integral diffuse back lighting.
Supplied with: mains lead and instruction manual.
SPECIFICATIONS
Temperature range |
-40 to +20°C |
Temperature stability |
±0.015 at -20°C (±0.01 at -40°C) |
Tube capacity |
up to 2 |
Bath Fluid |
Methyl Alcohol or Denatured Ethyl Alcohol |
Bath capacity |
2.9 litres |
Refrigerant |
R404A |
Voltage |
220/240V, 50/60Hz |
Power |
780W |
Size (HxWxD) |
55 x 49 x 53cm |
Weight |
52kg |
CCCN Code: |
Tariff 90278091 |
METHODS
ASTM D445-IP 71, ASTM D446, BIS IS 1448 Part 25, BS 188, BS 2000 Part 71, DIN 51 561, FTM 791 305, ISO 3104, ISO 3105, JIS K2283, NF T60-100
DETAILS OF METHODS
ASTM D445-IP 71 |
Standard test method for kinematic viscosity of transparent and opaque liquids (the calculation of dynamic viscosity) |
BS 188 |
Methods for determination of the viscosity of liquids |
BS 2000 Part 71 |
Kinematic Viscosity |
ISO 3104 |
Kinematic Viscosity of Transparent and Opaque Liquids and the Calculation of Dynamic Viscosity |
ASTM D446 |
Standard specifications and operating instructions for glass capillary kinematic viscometers |
ISO 3105 |
Glass capillary kinematic viscometers |
FTM 791 305 |
Measurement of kinematic viscosity is vitally important for numerous products, both during and after formation, to assess conformance to specification and suitability for use. |
JIS K2283 |
Viscosity by BS/IP Viscometer |
DIN 51 561 |
Viscosity |
BIS IS 1448 Part 25 |
Determination of Kinematic and Dynamic Viscosity |
NF T60-100 |
Kinematic Viscosity of Transparent and Opaque Liquids and the Calculation of Dynamic Viscosity |